 Cyclex D Sampling Guide
Slide1

CYCLEX D SAMPLING GUIDE

SAMPLING PROCEDURE

    1.Remove blue pin from the bottom of the cyclex-d.

    2.Simply connect one end of tubing (5 feet or less) to the bottom of the cyclex-d (No adapters are necessary).

    3.Connect the other end of tubing to a pump precalibrated to a flow rate of 20 lpm.

    4.Remove red cap from the top of the cyclex-d.

    5.Turn pump on and pull a sample for 10 minutes, depending on environment (See ``Sampling Intervals").

    6.When sampling is complete, replace blue pin and red cap to the bottom and top of the cyclex-d.

Sampling Intervals Sampling Time
Clean (Outdoor/Indoor) 10 minutes
High Activity (Indoor) 5-8 minutes
Inner Wall Samples 2-5 minutes